A validated model for correcting Confounding Factors in Corneal Densitometry
DOI:
https://doi.org/10.26754/jji-i3a.202613315Resumen
Confounding factors (brightness, ACD, CCT, tilt) introduce systematic bias in corneal densitometry measurements. A validated correction model, developed with synthetic data (n=1000) and validated clinically (n=132), enables accurate isolation of age-related corneal transparency changes, improving the reliability and clinical interpretation of densitometry assessments.
Descargas
Los datos de descarga aún no están disponibles.
Descargas
Publicado
2026-07-17
Número
Sección
Artículos (Ingeniería Biomédica)
Licencia
Derechos de autor 2026 Ana R. Arizcuren, Alejandra Consejo

Esta obra está bajo una licencia internacional Creative Commons Atribución-NoComercial 4.0.
Cómo citar
Arizcuren, A. R., & Consejo, A. (2026). A validated model for correcting Confounding Factors in Corneal Densitometry. Jornada De Jóvenes Investigadores Del I3A, 14. https://doi.org/10.26754/jji-i3a.202613315
Datos de los fondos
-
European Research Council
Números de la subvención 101162733